LYSO crystal calorimeter readout with silicon photomultipliers

Author:
A. Berra  V. Bonvicini  C. Cecchi  S. Germani  D. Guffanti  D. Lietti  P. Lubrano  E. Manoni  M. Prest  A. Rossi  E. Vallazza  


Journal:
Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment


Issue Date:
2014


Abstract(summary):

Abstract Large area Silicon PhotoMultipliers (SiPMs) are the new frontier of the development of readout systems for scintillating detectors. A SiPM consists of a matrix of parallel-connected silicon micropixels operating in limited Geiger–Muller avalanche mode, and thus working as independent photon counters with a very high gain (~10 6 ). This contribution presents the performance in terms of linearity and energy resolution of an electromagnetic homogeneous calorimeter composed of 9 ~ 18 X 0 LYSO crystals. The crystals were readout by 36 4×4 mm 2 SiPMs (4 for each crystal) produced by FBK-irst. This calorimeter was tested at the Beam Test Facility at the INFN laboratories in Frascati with a single- and multi-particle electron beam in the 100–500 MeV energy range.


Page:
248-248